Description


CONTEC NEW Touch Screen ECG Machine Electrocardiograph
 
Model:ECG600G


Introduction

ECG100G is a kind of single channel electrocardiograph, which features in  printing ECG waveform with thermal printing system, prompting for "Lead  off"
or "Lack of paper", multi-language interface, recording ECG  waveform with manual or auto mode, convenient to operate.


Features
1)Operation

  • 2.7" STN-LCD, accurate and clear ECG waveform and working state display, economical to save recording paper.
  • Touch keyboard, simple and convenient to operate.
2)Functions
  • Adopt digital signal processing technology to get high-quality ECG waveforms  via AC filter, baseline filter and EMG filter of ECG signal.
  • Real-time record single-channel ECG waveform and annotation(including lead mark,  sensitivity, paper speed and filter state, etc.).
  • Adopt high-resolution thermal printing system(8 dot/mm), no need for any adjustments. Recording frequency: up to 150Hz.
  • In AUTO mode, finish all operation by pressing the button once, which improves work efficiency.
  • Multi-language (Chinese, English, French, Italian, Turkish, Spanish and German) interface and report.
3)Power supply
  • AC/DC, built-in rechargeable lithium battery.
  • In optimal DC state, up to 7h standby time, continuous print up to 4h, record up to 160 ECG waveform.
Parameters
  • Working environment
    Temperature: 5℃~40℃
    Relative humidity: 25%~80%(non-condensing)
    Atmospheric pressure: 700hPa~1060hPa
  • Transport and storage environment
    Temperature: -40℃~+55℃
    Relative humidity: ≤95%
    Atmospheric pressure: 500hPa~1060hPa
  • Input mode: floating and defibrillation protection
  • Frequency response: 0.05Hz ~ 150Hz(-3dB~+0.4dB)
  • CMRR: >60dB, >100dB(add filter)
  • Time constant:≥ 3.2s
  • Patient leak current: <10µA
  • Calibration voltage: 1mV
  • Sensitivity: 5, 10, 20 mm/mV ±5%, standard sensitivity: 10mm/mV±2%
  • Noise level: ≤15µVp-p
  • Input circuit current: ≤50nA
  • Input impedance: ≥50MΩ
  • Sampling accuracy: 12-bit
  • Safety classification: class Ⅰ, type CF and defibrillation-proof applied part
  • Recording mode: thermal printing system
  • Paper size: 50mm(W)×20m(L)
  • Sampling frequency: 800Hz
  • Polarizing voltage: ±300mV
  • EMG interference filter: 35Hz(-3dB)
  • AC filter: 50Hz/60Hz(-20dB)
  • Paper speed: 25, 50mm/s,±5%
  • Power supply:
    AC: 100V~240V(50/60Hz)
    DC: 7.4V/2000mAh rechargeable lithium battery
  • Fuse specification: two AC time lag fuse(φ5×20mm), T1.6AL250V
  • Waterproof degree: IPX0
  • Working mode: continuous working
  • Dimension: 315mm(L)×215mm(W)×77mm(H)
  • Weight: 1.5Kg
Accessories
  • A lead cable
  • A limb electrode
  • A chest electrode
  • A thermal recording paper
  • A power cord
  • An earth wire
  • A user manual
Optional:
  • Bag
  • Conductive gel
Qualification Certificate

CE, FDA