Style name: We will inform you if the body cuff is wrapped perfectly.
General name: Automatic Electronic Blood Pressure Monitor (Classification: 167)
Product name: Omron upper arm type blood pressure monitor Contraindication / prohibition:
[Contraindication / Prohibition]
· Self-judgment of measurement result or treatment by self judgment
[It may lead to worsening of symptoms. ]
· Measurement with injuries or arms under treatment
[It may lead to worsening of symptoms. ]
· Measurement with arms during intravenous injection or transfusion
[It may lead to worsening of symptoms. ]
· Use in the vicinity of flammable, combustible gas
[It may cause ignition, ignition or explosion. ]
· Use beyond the useful life
[It may not be able to measure correctly. ]
· Use of unspecified number of people at medical institutions and public places
[It will cause accidents and troubles. ]
Item specification etc.:
1. Pressure
(1) Measurement range: 0 to 299 mmHg
(2) Scale: 1 mmHg
(3) Pressure: Automatic pressure method
(4) Exhaust: Automatic rapid exhaust system
2. Pulse
(1) Measurement range: 40 to 180 beats / minute
(2) Display unit: 1 beat / minute
(3) Measurement accuracy: Within ± 5% of reading value
3. Target arm circumference: 22 ~ 32 cm
4. Error in blood pressure measurement by clinical performance test:
Average error with auscultatory method; within ± 5 mmHg Standard deviation: within 8 mmHg
5. Error in cuff pressure indication: Within ± 3 mmHg
6. Rapid Exhaust: Time required to rapidly evacuate from 260 mmHg to 15 mmHg; 10 seconds or less
7. Stability of cuff pressure indication: Changes in display value of cuff pressure after simulation measurement of 10,000 cycles; 3 mmHg or less Caution on specifications:
1. If you have severe blood circulation disorder or blood disorder, please use under the guidance of a doctor.
2. When abnormality occurs, please remove the arm band from the arm.
3. Please do not use for infants or those who can not express their intention by themselves.
4. Do not press the arm band more than 300 mmHg.
5. Do not use for purposes other than blood pressure measurement.
6. Do not use anything other than a special armband.
7. Do not use mobile phones near the sphygmomanometer.
8. Do not disassemble or remodel the body or cuff.
9. Insert the + - pole of the battery correctly.
10. Please use the specified battery.
11. Remove the battery for a long period (over 3 months) when not in use.
Information on Manufacturers and Manufacturers:
Manufacturer selling agency: Omron Healthcare Co., Ltd. 53, Kinonobori Terado Town, Tominagaoka, Kyoto Prefecture 617-0002 Tel: 0120-30-6606
Effect / indication:
To measure systolic blood pressure and diastolic blood pressure non-invasively for health management.