Features
1)Operation
- 8" TFT color LCD, clear ECG waveform and working state display, economical to save recording paper.
- Operation mode: touch screen and function keys.
- Sync collection for 12-lead ECG, adopt digital signal processing technology to get high-quality ECG waveforms via AC filter, baseline filter and EMG filter of ECG signal.
- Sync display 3/6/12-lead ECG, print mode, sensitivity, paper speed and filter state, etc. on one screen, convenient to contrast.
- Multiply printing modes and formats, including manual, 12*1, 6*2+1(rhythm lead), 6*2, 3×4+2(rhythm lead), rhythm 12, rhythm 10, rhythm 8, rhythm 6, etc. Waveform length printed can be adjusted, and with timing print function, which meets different requirements.
- Paper speed: 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s.
- With the functions of auto-analysis and auto-diagnosis for routine ECG parameters, provide measurement results and auto-diagnosis conclusion for HR, P-R interval, P Duration, QRS Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+S(V1), etc. which reduces the doctor’s burden.
- Built-in memory stores up to 1000 cases, convenient for case review and statistic.
- Multi-language(Chinese, English, Turkish, Portuguese, German, Russian, Kazakhstan) interface and report.
- AC/DC, and in optimal DC state, built-in rechargeable lithium battery can continuously work for 10h, print for more than 3h and 300-ECG, which meets the requirements of visiting a patient at home and body examination.
- Working environment
Relative humidity: 20%~80%(non-condensing)
Atmospheric pressure: 700hPa~1060hPa
- Transport and storage environment
Relative humidity: ≤95%
Atmospheric pressure: 500hPa~1060hPa
- Input mode: floating and defibrillation protection
- Frequency response: 0.05Hz ~ 150Hz(-3dB~+0.4dB)
- CMRR: >60dB, >100dB(add filter)
- Time constant:> 3.2s
- Patient leak current: <10µA
- Calibration voltage: 1mV
- Sensitivity: 2.5, 5, 10, 20, 40 mm/mV ±5 %, standard sensitivity: 10mm/mV±2%
- Noise level: ≤15µVp-p
- Input circuit current: ≤50nA
- Input impedance: ≥50MΩ(10Hz)
- Sampling accuracy: 12-bit
- Safety classification: class Ⅰ, type CF and defibrillation-proof applied part
- Recording mode: thermal printing system
- Paper size: 210mm(W)×20m(L)
- Sampling frequency: 1000Hz
- Polarizing voltage: ±500mV
- EMG interference filter: 25Hz/35Hz (-3dB)
- AC filter: 50Hz/60Hz(-20dB)
- Paper speed:
Rhythm record: 25, 50 mm/s, ±5%
Manual record: 5, 6.25, 10, 12.5, 25, 50 mm/s, ±5%
- Auto record: set record according to auto record format and mode, automatically switching leads, measuring and analyzing.
- Rhythm record: set record according to rhythm record format and mode, automatically measuring and analyzing.
- Manual record: set record according to record format, manually switching leads.
- Measurement parameters: HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+S(V1).
- Power supply:
DC: 7.4V/2000mAh rechargeable lithium battery
- Fuse specification: two AC time lag fuse(φ5×20mm), T1.6AL250V
- Waterproof degree: IPX0
- Working mode: continuous working
- Dimension: 340mm(L)×320mm(W)×85mm(H)
- Weight: about 3.2Kg
- A lead cable
- A limb electrode
- A chest electrode
- A thermal recording paper
- A power cord
- An earth wire
- A user manual
- ECG-Sync software+USB cable
- Disposable ECG electrode
- Bag
- Conductive paste
- USB cable
- 3G card
- Wi-Fi card
- Remote Operation Guide
CE, FDA