LCD Digital Pen Type PH Meter Pocket Pink PH Pen Acidimeter High Precision Waterproof

Instrument Tester Range 0.00-14.00pH New Arrival

Caution:
 Please read the manual carefully before using the meter.
1. The glass electrode at the bottom of the pen is fragile, please use it carefully after taking off the protection cap.
Any damage may cause the invalid of the electrode.
2. The duration period of the glass electrode is one year from the date of purchase whether you use it or not,
please change the electrode in time.
3. The electrode can not be used under the dry condition, always soak the electrode into the distilled water or
purified water for 5 – 30 minutes to activate before use.
Product Name: CT-6022-PH (pen) acidimeter
2 Products:
   CT-6022 the PH pen acidimeter high-performance CPU chip, high-precision AD converter technology and SMT technology is a versatile, intelligent instrument. Storage, computing and Compensation Determination of PH value parameters. Clear display, simple operation and accurate test performance so that it has a good price. It can be widely used in the chemical industry, metallurgy, environmental protection, pharmaceutical, biochemical, food and tap water solution PH value monitoring. Field PH value test.
3 acidimeter Features:
     The instrument housing waterproof design of acidimeter, LCD display. Small size, portable, measuring at any time more convenient to use. Low power consumption, high accuracy, wide measurement range, the electrode can be replaced.
4 Acidimeter technical indicators:
Measuring range:0.00-14.00pH
Accuracy:±0.02pH After calibration
Resolution:0.01pH  
Automatic temperature compensation:0C-50C
Operating Temperature:0C-50C
Correction:3 point automatic calibration (4.01; 6.86;9.18)
Automatic shutdown:10 minute auto shut off
Low voltage indication
Have Automatic conversion of C / F
Have Hold Function
Have Low voltage indication
Have Display
Dual display of pH and temperature
Battery:6V (4 * AG13 button batteries)(Not Included)
CBM:188mm×38mm
Weight:82g
Calibration
1. Prepare 4.01, 6.86 and 9.18 pH buffers, Use 6.86 pH buffer for the mid-range buffer first. The pH values for the
buffers are given for 25 °C. If the sample temperature is not 25 °C, the pH values displayed for the buffers will
reflect the correct pH value for the sample temperature. If the electrode is dry, submerge it in distilled water for 10
minutes before calibration. Ensure that the calibrating buffers are fresh.
2. Press ON/OFF to turn the meter on, submerge the probe in the buffer while stir gently. Then keep it still until a
stable reading is reached. Press CAL for 3 seconds until the Text `CAL` appears. Then release CAL, the meter will
identify the current buffer value automatically, and display 6.86 in the LCD. The result will be saved while the text
`SA` displayed after 2 seconds. The meter will back to measuring mode after 1 second while text `End` showed.
3. If text `End` is showed after press CAL, it means the calibration buffer is not fresh or the probe is aging.
4. Do not take out the probe from the buffer until text `End` is showed in the LCD.
5. Same steps as calibrating pH 4.01 and 9.18.

Shipment:

When you place an order,we will send the items within 3 days once your payment is completed.

We do not guarantee delivery time on all international shipments due to differences in customs clearing times in individual countries, which may affect how quickly your product is inspected.

Please note that buyers are responsible for all additional customs fees, brokerage, fees, duties, and taxes for importation into your country. These additional fees may be collected at time of delivery. We will not refund shipping charges for refused shipments.

Warranty & Maintenance:

We offer the 15 days free replacement.  Buyers may request for a replacement within 7 days from your receipt of the product. The buyer should return the product in the original conditions to us and should bear the shipping costs for return.

Feedback:

Your satisfaction and positive feedback is very important to us.  Please leave positive feedback and 5 stars if you are satisfied with our items and services.

The Most Important!!!

If you have any problems with our items or services, please feel free to contact us first before you leave negative feedback.